Microliter Flow Cell PEEK 0.5 mm

Ultra-compact fiber-optic flow cell for UV/VIS and NIR applications with a 0.5 mm optical path length, 5 µl internal volume, and a broad spectral range from 200 to 2400 nm. Thanks to SMA-905 fiber connectors and 10-32 UNF fluid connections, it can be easily integrated into HPLC, microfluidic, FIA, and process analysis systems.

Microliter Flow Cell PEEK 0.5 mm

This compact, biocompatible fiber-optic HPLC flow cell combines highly durable PEEK with high-quality quartz glass and enables precise measurements in the UV/VIS and NIR spectral ranges. The short optical path length prevents signal clipping and makes the flow cell ideal for high-concentration samples without dilution. Thanks to its high pressure resistance and robust design, it is ideally suited for demanding applications in UHPLC, HPLC, FIA, GFIA, as well as for inline and process analytics.

 

Product features

  • 0.5 mm optical path length for the analysis of highly concentrated samples
  • Optimized for HPLC, UHPLC, FIA, and microfluidic applications
  • Extremely small internal volume of just 5 µl to minimize dead volume and peak broadening
  • Very low dispersion volume for maximum chromatographic resolution
  • Wide spectral range from 200 to 2400 nm for UV, VIS, and NIR measurements
  • SMA-905 fiber connectors, optimized for optical fibers with a 600 µm core diameter
  • Chemically resistant and biocompatible materials made of PEEK and quartz glass
  • Pressure rating: 50 bar (test pressure: 60 bar)
  • Compatible with 1/16″ tubing systems and 10-32 UNF fittings
  • Compact design for easy integration into analytical and process analytical systems

 

Technical data

Category Specification
Path length 0.5 mm ± 0.03 mm
Wavelength range1 200 nm – 2,400 nm
Transmission 1, 2 > 40%
Drift (ASTM E1657) < 1 mAU/h at 254 nm and 1 mL/min
Internal volume 5 µl
Flow rate 0–20 ml/min, typical: 0.05–1 ml/min
Temperature range 5–80 °C
Pressure rating3 50 bar (60 bar test pressure)
Fluid Connection 10‑32 UNF coned fitting for PEEK 1/16″ & 1/32″ tubing
Optical Connector SMA 905, optionally available with anti-twist feature, optimized for 600 µm
Windows Quartz glass
Wetted Materials PEEK, quartz glass
Dimensions Φ = 30 mm, l = 78 mm

1 Flow cellfilled with ultrapure water (18 MOhm, TOC < 10 ppb)
2 Transmission is measured using two directly coupled fibers with a core diameter of 600 µm in the wavelength range between 200 nm and 1000 nm as a reference, and filled as described in  1
3. Higher pressure ratings available upon request

 

Areas of application

  • HPLC and UHPLC Detection
  • UV/VIS and NIR In-Line Analysis
  • Flow Injection Analysis (FIA) and Gas-Segmented Flow Injection Analysis (GFIA)
  • Microfluidic Measurement and Analysis Systems
  • Process Analytics (PAT)
  • Liquid and Media Monitoring
  • Concentration and Quality Monitoring
  • Research and development
  • Sensor Integration and OEM Applications